Directions for Using the Rossum Sand Tester
The Rossum Sand Tester is used to measure sand content of pumped water. These directions will allow the user to calculate this measurement in units represented as parts per million by volume.
The Rossum Sand Tester can be used to measure sand content in any water system. However, the water sample for sand determination must be carefully obtained. A sample collected from a horizontal pipe at low flow in unacceptable. The sample should be obtained from a vertical or horizontal pipe with highly turbulent flow. Samples taken immediately downstream from elbows, tees, and other fittings that create turbulence are generally satisfactory.
Procedure
1. Install the Rossum Sand Tester as shown in drawing 1
. The inlet should be located on the horizontal centerline of the discharge pipe and be as close to the discharge head or other turbulent flow area as possible.
2. Completely open the inlet valve to the tester. Adjust the outlet valve to 1/2 gpm (this will fill a quart container in 30 seconds or a gallon container in 2 minutes).
3. Close the inlet valve, remove, clean, and replace the Pyrex tube.
4. When ready to start the test, completely open the inlet valve and note the time.
5. Periodically record the accumulated sand and time. It is important to note, the most common mistake made when doing the final calculation is dividing total
accumulation of sand by total time. This provides only the average sand content. Instead, it is necessary to divide the incremental sand increase by the
incremental time increase. This will provide the sand production at specific time
ie. sand production 15 minutes after the start of the test.
6. Check the flow rate through the tester during each run. If the flow rate is not 1/2
gpm, repeat the test.
7. Record the pump discharge rate (gpm) during each sand test and whether any
surging was done prior to the test.